Exploring the Voltage-Clamp Membrane Test

By modeling a voltage-clamp amplifier, patch pipette, and cell membrane as a circuit using free circuit simulation software, I was able to create a virtual patch-clamp electrophysiology workstation and challenge model neurons with advanced voltage-clamp protocols. By modeling neurons with known properties and simulating experimental membrane test protocols, I can write membrane test analysis software and confirm its accuracy by comparing my calculated membrane measurements to the values in the original model. A strong advantage of this method (compared to using physical model cells) is that I can easily change values of any individual component to assess how it affects the accuracy of my analytical methods.

Instead of modeling a neuron, I modeled the whole patch-clamp system: the amplifier (with feedback and output filtering), pipette (with an imperfect seal, series resistance, and capacitance), and cell (with membrane resistance, capacitance, and a resting potential). After experimenting with this model for a while I realized that advanced topics (like pipette capacitance compensation, series resistance compensation, and amplifier feedback resistance) become much easier to understand when they are represented as components in a circuit with values that can be adjusted to see how the voltage-clamp trace is affected. Many components of the full model can be eliminated to generate ideal traces, and all models, diagrams, and code shown here can be downloaded from my membrane test repository on GitHub.

Circuit Components

Cell

Pipette

Amplifier

Modeling a Patch-Clamp Experiment in LTSpice

LTSpice is a free analog circuit simulator by Analog Devices. I enjoy using this program, but only because I’m used to it. For anyone trying to use it for the first time, I’m sorry. Watch a YouTube tutorial to learn how to get up and running with it. Models used in this project are on GitHub if you wish to simulate them yourself.

This circuit simulates a voltage clamp membrane test (square pulses, ±5mV, 50% duty, 20 Hz) delivered through a patch pipette (with no pipette capacitance), a 1GΩ seal, 15 MΩ access resistance, in whole-cell configuration with a neuron resting at -70 mV with 500 MΩ membrane resistance and 150 pF capacitance. The Bessel filter is hooked-up through a unity gain op-amp so it can be optionally probed without affecting the primary amplifier. It’s configured to serve as a low-pass filter with a cut-off frequency of 2 kHz.

Simulating a Membrane Test

The simulated membrane test shows a typical voltage-clamp trace (green) which is interesting to compare to the command voltage (red) and the actual voltage inside the cell (blue). Note that although the hardware low-pass filter is connected, the green trace is the current passing through the feedback resistor (Rf). A benefit of this simulation is that we can probe anywhere, and being able to see how the cell’s actual voltage differs from the target voltage is enlightening.

If your clamp voltage does not have sharp transitions, manually define rise and fall times as non-zero values in the voltage pulse configuration options. Not doing this was a huge trap I fell into. If the rise time and fall time is left at 0, LTSpice will invent a time for you which defaults to 10%! This slow rise and fall of the clamp voltage pulses was greatly distorting the peaks of my membrane test, impairing calculation of I0, and throwing off my results. When using the PULSE voltage source set the rise and fall times to 1p (1 picosecond) for ideally sharp edges.

If saving simulation data consider defining the maximum time step. Leaving this blank is typically fine for inspecting the circuit within LTSpice, but if you intend to save .raw simulation files and analyze them later with Python (especially when using interpolation to simulate a regular sample rate) define the time step to be a very small number before running the simulation.

Low-Pass Filtering

Let’s compare the output of the amplifier before and after low-pass filtering. You can see that the Bessel filter takes the edge off the sharp transient and changes the shape of the curve for several milliseconds. This is an important consideration for analytical procedures which seek to measure the time constant of the decay slope, but I’ll leave that discussion for another article.

Calculate Clamp Current from Amplifier Output Voltage

Patch-clamp systems use a digital-to-analog converter which measures voltage coming out of the amplifier to infer the current being delivered into the pipette. In other words, the magic ability LTSpice gives us to probe current passing through any resistor in the circuit isn’t a thing in real life. Instead, we have to use Ohm’s law to calculate it as the ratio of voltage and feedback resistance.

Let’s calculate the current flowing into the pipette at the start of this trace when the amplifier’s output voltage is -192 mV and our command potential is -75 mV:

V = I * R
I = V / R
I = (Vout - Vcmd) / Rf
I = ((-192e-3 V) - (-75e-3 V)) / 500e6 Ω
I = -234 pA

Notice I use math to get the difference of Vout and Vcmd, but in practice this is done at the circuit level using a differential amplifier instead of a unity gain op-amp like I modeled here for simplicity.

Amplifier Feedback Capacitance

Let’s further explore this circuit by adding pipette capacitance. I set Cp to 100 pF (I know this is a large value) and observed strong oscillation at clamp voltage transitions. This trace shows voltage probed at the output of the Bessel filter.

A small amount of feedback capacitance reduced this oscillation. The capacitor Cf placed across Rf serves as an RC low-pass filter to tame the amplifier’s feedback. Applying too much capacitance slows the amplifier’s response unacceptably. It was impressive to see how little feedback capacitance was required to change the shape of the curve. In practice parasitic capacitance likely makes design of patch-clamp amplifier headstages very challenging. Experimenting with different values of Cp and Cf is an interesting experience. Here setting Cp to 1 pF largely solves the oscillation issue, but its low-pass property reduces the peaks of the capacitive transients.

Two-Electrode Giant Squid Axon Model

I created another model to simulate a giant squid axon studied with a two-electrode system. It’s not particularly useful other than as a thought exercise. By clamping between two different voltages you can measure the difference in current passing through the stimulation resistor to estimate the neuron’s membrane resistance. This model is on GitHub too if you want to change some of the parameters and see how it affects the trace.

Let’s calculate the squid axon’s membrane resistance from the simulation data just by eyeballing the trace.

ΔV = (-65 mV) - (-75 mV) = 10 mV <-- Δ command voltage
ΔI = (5 µA) - (-5 µA) = 10 µA <-- Δ amplifier current
V = I * R
ΔV = ΔI * Rm
Rm = ΔV / ΔI
Rm = 10e-3 V / 10e-6 A
Rm = 1kΩ <-- calculated membrane resistance

Load LTSpice Simulation Data with Python

LTSpice simulation data is saved in .raw files can be read analyzed with Python allowing you to leverage modern tools like numpy, scipy, and matplotlib to further explore the ins and outs of your circuit. I’ll discuss membrane test calculations in a future post. Today let’s focus on simply getting these data from LTSpice into Python. Simulation data and full Python code is on GitHub. Here we’ll analyze the .raw file generated by the whole-cell circuit model above.

# read data from the LTSpice .raw file
import ltspice
l = ltspice.Ltspice("voltage-clamp-simple.raw")
l.parse()

# obtain data by its identifier and scale it as desired
times = l.getTime() * 1e3 # ms
Vcell = l.getData('V(n003)') * 1e3  # mV
Vcommand = l.getData('V(vcmd)') * 1e3  # mV
Iclamp = l.getData('I(Rf)') * 1e12  # pA

# plot scaled simulation data
import matplotlib.pyplot as plt

ax1 = plt.subplot(211)
plt.grid(ls='--', alpha=.5)
plt.plot(times, Iclamp, 'r-')
plt.ylabel("Current (pA)")

plt.subplot(212, sharex=ax1)
plt.grid(ls='--', alpha=.5)
plt.plot(times, Vcell, label="Cell")
plt.plot(times, Vcommand, label="Clamp")
plt.ylabel("Potential (mV)")
plt.xlabel("Time (milliseconds)")
plt.legend()

plt.margins(0, .1)
plt.tight_layout()
plt.show()

LTSpice simulation data points are not evenly spaced in time and may require interpolation to produce data similar to an actual recording which samples data at a regular rate. This topic will be covered in more detail in a later post.

Membrane Test Analysis

Let’s create an ideal circuit, simulate a membrane test, then analyze the data to see if we can derive original values for access resistance (Ra), cell capacitance (Cm), and membrane resistance (Rm). I’ll eliminate little tweaks like seal resistance, pipette capacitance, and hardware filtering, and proceed with a simple case voltage clamp mode.

⚠️ WARNING: LTSpice voltage sources have a non-negligible conductance by default, so if you use a voltage source at the base of Rm without a defined resistance you’ll have erroneous steady state current readings. Prevent this by defining series resistance to a near infinite value instead of leaving it blank.

Now let’s run the simulation and save the output…

I created a diagram to make it easier to refer to components of the membrane test:

Think conceptually about what’s happening here: When the command voltage abruptly changes, Vcell and Vcommand are very different, so the voltage-clamp amplifier delivers a large amount of current right after this transition. The peak current (Ipeak) occurs at time zero relative to the transition. The current change between the previous steady-state current (Iprev) and the peak current (Ipeak) is only limited by Ra (since Cm only comes in to play after time passes). Let’s call this maximum current change Id. With more time the current charges Cm, raising the Vcell toward (Vcommand) at a rate described by TauClamp. As Vcell approaches Vcommand the amplifier delivers less current. Altogether, amplifier current can be approximated by an exponential decay function:

It = Id * exp(-t / τclamp) + Iss

Analyze the Capacitive Transient

The speed at which Vcell changes in response to current delivered through the pipette is a property of resistance (Ra) and capacitance (Cm). By studying this curve, we can calculate both. Let’s start by isolating one curve. We start by isolating individual capacitive transients:

Fit each curve to a single exponential function. I’ll gloss over how to do this because it is different for every programming language and analysis software. See my Exponential Fit with Python for details. Basically you’ll fit a curve which has 3 parameters: m, tau, and b. You may wish to change the sign of tau depending on the orientation of the curve you are fitting. If your signal is low-pass filtered you may want to fit a portion of the curve avoiding the fastest (most distorted) portion near the peak. If you want to follow along, code for this project is on GitHub.

It = m * exp(-t / tau) + b

These are the values I obtained by fitting the curve above:

m = 667.070
tau = 2.250
b = -129.996

Meaning the curve could be modeled by the equation:

I = 667.070 * exp(-t / 2.250) -129.996

From these values we can calculate the rest:

We now have:

Iss: -129.996 pA
Iprev: -150.015 pA
Idss: 20.019 pA
Ipeak: 537.074 pA
Id: 687.089 pA
dV: 10 mV
TauClamp: 2.250 ms

Calculate Ra

At time zero, access resistance is the thing limiting our ability to deliver current (Id) to a known ΔV (10 mV). Therefore we can calculate Ra using Ohm’s law:

V = I * R
ΔV = ΔI * R
R = ΔV / ΔI
Ra = dV / Id
Ra = 10e-3 V / 687.089e-12 A
Ra = 14.554 MΩ <-- pretty close to our model 15 MΩ

For now let’s call this Ra, but note that this is technically Ra mixed with a small leakage conductance due to Rm. Since Ra is so much smaller than Rm this small conductance doesn’t affect our measurement much. Accuracy of this value will be improved when we apply leak current correction described later on this page.

Calculate Rm

Now that we know Ra, we can revisit the idea that the difference between this steady state current (Iss) and the last one (Iprev) is limited by the sum of Rm and Ra. let’s use this to calculate Rm using Ohm’s law:

V = I * R
I = V / R
ΔI = ΔV / R
R * ΔI = ΔV
(Ra + Rm) * ΔI = ΔV
Ra * ΔI + Rm * ΔI = ΔV
Rm * ΔI = ΔV - Ra * ΔI
Rm = (ΔV - Ra * ΔI) / ΔI
Rm = (dV - Ra * Idss) / Idss
Rm = (10e-3 V - (14.554e6 Ω * 20.019e-12 A)) / 20.019e-12 A
Rm = 485 MΩ <-- pretty close to our model 500 MΩ

Accuracy of this value will be improved when we apply leak current correction described later on this page.

Calculate Cm from Ra, Rm, and Tau

When we raise the cell’s voltage (Vm) by delivering current through the pipette (Ra), some current escapes through Rm. From the cell’s perspective when we charge it though, Ra and Rm are in parallel.

tau = R * C
C = tau / R
Cm = tau / (1/(1/Ra + 1/Rm))
Cm = 2.250e-3 sec / (1/(1/14.554e6 Ω + 1/485e6 Ω))
Cm = 159 pF <-- pretty close to our model 150 pF

Accuracy of this value will be improved when we apply leak current correction described later on this page.

Calculate Cm from the Area Under the Curve

Cell capacitance can alternatively be estimated by measuring the area under the capacitive transient. This method is frequently used historically, and it is simpler and faster than the method described above because it does not require curve fitting. Each method has its pros and cons (e.g., sensitivity to access resistance, hardware filtering, or resilience in the presence of noise or spontaneous synaptic currents). Rather than compare and contrast the two methods, I’ll simply describe the theory underlying how to perform this measurement.

After an abrupt voltage transition, all current delivered above the steady state current level goes toward charging the cell, so by integrating this current over time we can calculate how much charge (Q) was delivered. I’ll describe this measurement as area under the curve (AUC). When summing these data points yourself be sure to remember to subtract steady state current and divide by the sample rate. Code for this example is on GitHub.

Charge is measured in Coulombs. Area under the curve is 1515.412 pA*ms, but recall that a femtocoulomb is 1pA times 1ms, so it’s more reasonable to describe the AUC as 1515.412 fC. This is the charge required to raise cell’s capacitance (Cm) by dV. The relationship is described by:

Q = C * ΔV
C = Q / ΔV
Cm = AUC / ΔV
Cm = 1515.412e-15 C / 10e-3 V
Cm = 1515.412e-15 C / 10e-3 V
Cm = 151.541 pF <-- pretty close to our model 150 pF

This value is pretty close to what we expect, and I think its accuracy in this case is largely due to the fact that we simulated an ideal unfiltered voltage clamp trace with no noise. Its under-estimation is probably due to the fact that a longer period wasn’t used for the integration (which may have been useful for this noise-free simulation, but would not be useful in real-world data). Additional simulation experiments with different combinations of noise and hardware filtering would be an interesting way to determine which methods are most affected by which conditions. Either way, this quick and dirty estimation of whole-cell capacitance did the trick in our model cell.

Correcting for Leak Current

Why weren’t our measurements exact? Rm leaks a small amount of the Id current that passes through Ra to charge Cm. If you calculate the parallel combined resistance of Ra and Rm you get 14.56 MΩ which is pretty much exactly what we measured in our first step and simply called Ra at the time. Now that we know the value of both resistances we can calculate a correction factor as the ratio of Ra to Rm and multiply it by both of our resistances. Cm can be corrected by dividing it by the square of this ratio.

correction = 1 + Ra / Rm
correction = 1 + 14.554 MΩ / 484.96 MΩ
correction = 1.03

Ra = Ra * correction
Rm = Rm * correction
Cm = Cm / (correction^2)
Metric Model Measured Corrected Error
Ra 15 MΩ 14.55 MΩ 14.99 MΩ <1%
Rm 500 MΩ 484.96 MΩ 499.51 MΩ <1%
Cm (fit) 150 pF 159.20 pF 150.06 pF <1%

This correction is simple and works well when Ra/Rm is small. It’s worth noting that an alternative to this correction is to solve for Ra and Rm simultaneously. The Membrane Test Algorithms used by pCLAMP calculate Ra this way, solving the following equation iteratively using the Newton-Raphson method:

Ra^2 - Ra * Rt + Rt * (Tau/Cm) = 0

Overall the values I calculated are within a few percent of expectations, and I’m satisfied with the calculation strategy summarized here. I am also impressed with what we were able to achieve by modeling a voltage-clamped neuron using a free circuit simulator!

Use a Voltage-Clamp Ramp to Measure Cm

It’s possible to simulate a voltage-clamp ramp and analyze that trace to accurately measure cell capacitance. A strong advantage of this method is that it does not depend on Ra. Let’s start by simulating a 10 mV ramp over 100 ms (50 ms down, 50 ms up). When we simulate this with LTSpice and plot it with Python (screenshots, data, and code is on GitHub) we find that cell voltage lags slightly behind the clamp voltage.

During voltage-clamp ramps Vm lags behind the command voltage because charging Cm is limited by Ra. If we measure the difference in this lag between descending and ascending ramps, we can estimate Cm in a way that is insensitive to Ra. Stated another way, Ra only affects abrupt changes in charging rate. Once the cell is charging at a steady rate, that rate of charge is largely unaffected by Ra because the stable charging current is already increased to counteract the previous effect Ra. Stated visually, Ra only affects the rate of charging at the corners of the V. Therefore, let’s proceed ignoring the corners of the V and focus on the middle of each slope where the charging rate is stable (and effect of Ra is negligible).

Analysis is achieved by comparing the falling current to the rising current. We start separately isolating the falling and rising traces, then reverse one of them and plot the two on top of each other. The left and right edges of this plot represent edges of ramps where the system is still stabilizing to compensate for Ra, so let’s ignore that part and focus on the middle where the charging rate is stable. We can measure the current lag as half of the mean difference of the two traces. Together with the rate of charge (the rate of the command voltage change) we have everything we need to calculate Cm.

dI = dQ / dt
dI = Cm * dV / dt
Cm = dI / (dV / dT)
Cm = (59.997e-12 A / 2) / (10e-3 V / 50e-3 sec) <-- 10 mV over 50 ms
Cm = 149.993 pF <-- Our model is 150 pF

This is a fantastic result! The error we do get is probably the result of a single point of interpolation error while converting the unevenly spaced simulation data to an evenly-spaced array simulating a 20 kHz signal. In this ideal simulation this method of calculating Cm appears perfect, but in practice it is highly sensitive to sporadic noise that is not normally distributed (like synaptic currents). If used in the real world each ramp should be repeated many times, and only the quietest sweeps (with the lowest variance in the difference between rising and falling currents) should be used for analysis. However, this is not too inconvenient because this protocol is so fast (10 repetitions per second).

Summary

This page described how to model voltage-clamp membrane test sweeps and analyze them to calculate Ra, Cm, and Rm. We validated our calculations were accurate by matching our calculated values to the ones used to define the simulation. We also explored measuring the area under the curve and using voltage-clamp ramps as alternative methods for determining Cm. There are a lot of experiments that could be done to characterize the relationship of noise, hardware filtering, and cell properties on the accuracy of these calculations. For now though, I’m satisfied with what we were able to achieve with free circuit simulation software and basic analysis with Python. Code for this project is on GitHub.

Metric Model Calculated Error
Ra 15 MΩ 14.99 MΩ <1%
Rm 500 MΩ 499.51 MΩ <1%
Cm (fit) 150 pF 150.06 pF <1%
Cm (auc) 150 pF 151.541 pF ~1%
Cm (ramp) 150 pF 149.993 pF <.01%

Resources