Confocal Microscope
How to use the Nikon C2 confocal microscope to image slide-mounted fluorescent samples
⚠
Warning:
This page may contain laboratory-specific information
Nikon C2 Confocal Microscope
- Brochure (official)
- NIS-Elements Instructions (official)
- Operating Notes (University of Minnesota)
- Training Guide (UT Dental)
- NIS-Elements Basics (YouTube)
- Imaging a Z-Stack (YouTube)
- Ramamurthy lab
- Getting Started
- Operating the Microscope Visually
- ??? (YouTube)
- ??? (YouTube)
- ??? (YouTube)
- ??? (YouTube)
- How to clean oil from the objectives
Startup
- do not turn on nikon white light source
- turn on sola fluorescent light source using power button on front
- Turn on Nikon C2 using power button on front
- Turn on laser switch on the back left by the power cord and turn key to ON
- Turn on proscan by flipping the switch on the back right DOWN
- open NIS elements software (select C2 confocal on startup)
Seeing fluorescent samples through the eyepiece
- press “light” button to enable fluroescence light source
- set path select knob on right of scope to “BINO”
- rotate filter cubes to 2(UV/blue), 3 (blue/green) or 4 (green/red)
- open the shutter on the front of the scope
- use the coarse focus knob on the scope
- use the joystick for fine focus and XY translation
Laser scanning confocal imaging
- press “light” button to disable fluroescence light source
- set path select knob on right of scope to “F”
- open the shutter on the front of the scope
- press “live” in the software